ALBERT

All Library Books, journals and Electronic Records Telegrafenberg

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
Filter
  • 1
    Electronic Resource
    Electronic Resource
    Oxford, UK : Blackwell Publishing Ltd
    Annals of the New York Academy of Sciences 523 (1988), S. 0 
    ISSN: 1749-6632
    Source: Blackwell Publishing Journal Backfiles 1879-2005
    Topics: Natural Sciences in General
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 2
    Electronic Resource
    Electronic Resource
    Oxford, UK : Blackwell Publishing Ltd
    Annals of the New York Academy of Sciences 523 (1988), S. 0 
    ISSN: 1749-6632
    Source: Blackwell Publishing Journal Backfiles 1879-2005
    Topics: Natural Sciences in General
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 3
    Electronic Resource
    Electronic Resource
    Hoboken, NJ : Wiley-Blackwell
    Journal of Biomedical Materials Research 22 (1988), S. 1203-1232 
    ISSN: 0021-9304
    Keywords: Chemistry ; Polymer and Materials Science
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Medicine , Technology
    Notes: The purpose of this article was to review the laboratory and clinical performances since 1970 of a total hip prosthesis using alumina-alumina combination. The chemical and physical properties of dense alumina ceramic were studied in relation to biocompatibility, mechanical strength, and surface properties. Through the examination of 35 retrieved implants, it was found that the long-term success of alumina-alumina total hip replacement depends on both the ceramic microstructure (small grain size with uniform distribution, minimum porosity, absence of inclusions) and implant geometry (sphericity deviation ±1 μm, radius tolerance between components 7-10 μm). Alumina component wear and fractures have disappeared with the use of high-performance materials and severe manufacturing quality control. Examination of human biopsies from well-fixed prostheses showed that alumina particles deposits increase with time with only a low-grade macrophagic reaction. When loosening occurred, an inflammatory reaction appeared; this reaction was less striking than with loose metal-polyethylene prostheses, however. The long-term behavior of cementless alumina cup fixation depends upon initial positioning and stability; survivorship analysis of the cemented ceramic cups showed an 88% survival probability after 8 years with a 1.6% average annual probability of revision. The percentage of surviving was 100% after 8 years in patients who were less than 50 years old. Aseptic loosenings occurring at the cup-cement interface were assumed to be related to stress protection secondary to the high rigidity of the ceramic leading to a weakening of the spongious bone supporting the cement mantle. Good bone stock quality as well as high-quality ceramic appear to be the prerequisites for durable fixation of alumina sockets.
    Additional Material: 18 Ill.
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 4
    Electronic Resource
    Electronic Resource
    Hoboken, NJ : Wiley-Blackwell
    Journal of Biomedical Materials Research 23 (1989), S. 299-310 
    ISSN: 0021-9304
    Keywords: Chemistry ; Polymer and Materials Science
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Medicine , Technology
    Notes: The wear behavior of hip prostheses alumina components was studied on 20 retrieved implants removed for socket loosening. The observed wear takes two forms: a running-in of the mating surfaces and a gross disruption of the surfaces. The running-in is very shallow, in the order of 0.2 m̈m after 8 years of implantation. This is observed on the load-bearing areas of the components. Gross wear is not material-related but had always been associated to a high contact pressure between the edge of the socket and the femoral head. Such a high contact pressure occurs after the socket had loosened and then tilted or if it was initially positioned too vertically. The long-term success of alumina - alumina THP components depends essentially upon biomechanical considerations and upon the quality of the material. The area of contact between the head and the socket should be large enough to minimize the contact stresses. This can only be achieved by a good positioning of the socket and by a close tolerance fit of the alumina components; any deviation from sphericity should also be avoided. The normal wear behavior of alumina is a running in process leading to an average wear rate of 0.025 m̈m/year as compared to 100 m̈m/year for UHMWPE on metallic heads.
    Additional Material: 7 Ill.
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...