Publication Date:
2006-02-14
Description:
Digitizing flatfield images produces conditions in the Photometric Data System PDS which cause the measured density to drift by as much as .1 DN during a 10 minute interval. The drift occurs when the PDS, set up in equilibrium at fog level, subsequently scans a reasonably dense region for periods of longer than a few minutes. The drift is manifested primarily as a positive shift in density that is approximately the same for all densities. If the fog level is assumed to be in fact constant and is monitored during scans of flat fields, the PDS drift may be removed by subtracting the difference between the observed fog level and its assumed constant value for each pixel. This function is then smoothed and subtracted, as a function of scan line, from the measured density. The fog level is then adjusted to a standard value by adding a constant. The result is a flattened scan with PDS drift removed to the accuracy within which the fog level drift matches the drift at other levels.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Astron. Microdensitometry Conf.; p 59-69
Format:
text
Permalink