Publication Date:
2016-07-16
Description:
We describe a prototype γ-ray detector based on a metallic magnetic calorimeter with an energy resolution of 46 eV at 60 keV and a reproducible response function that follows a simple second-order polynomial. The simple detector calibration allows adding high-resolution spectra from different pixels and different cool-downs without loss in energy resolution to determine γ-ray centroids with high accuracy. As an example of an application in nuclear safeguards enabled by such a γ-ray detector, we discuss the non-destructive assay of 242 Pu in a mixed-isotope Pu sample.
Print ISSN:
0003-6951
Electronic ISSN:
1077-3118
Topics:
Physics
Permalink