    Publication Date: 2011-08-24
    Description: A cryogenic blackbody radiation source for providing target temperatures between room temperature and -100 C was designed, fabricated, and tested. This blackbody radiation source is intended for calibration of radiometric infrared imaging systems down to -100 C. The construction and operating principles of this apparatus are described. The long term and short term stability of the radiant source have been investigated. The emittance of the source and the source radiant uniformity have been addressed.
    Keywords: INSTRUMENTATION AND PHOTOGRAPHY
    Type: In: International Instrumentation Symposium, 38th, Las Vegas, NV, Apr. 26-30, 1992, Proceedings (A93-37851 15-35); p. 273-285.
    Format: text
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 2
    Publication Date: 2011-08-24
    Description: Mineral-insulated metal-sheathed (MIMS) thermocouple assemblies are used throughout industry and research facilities as a method of temperature measurement where requirements for either harsh environmental conditions exist, or where rigidity of the measurement probe is required. Seebeck inhomogeneity is the abnormal variation of the Seebeck coefficient from point to point in a material. It is not disclosed in conventional calibration. A standardized method of measuring thermoelectric inhomogeneity along the thermocouple probe length is not available. Therefore, calibration for sheathed probes normally does not include testing of probe inhomogeneity. The measurement accuracy would be severely impacted if significant inhomogeneity and a temperature gradient were present in the same region of the probe. A computer-controlled system for determining inhomogeneities was designed, fabricated, and tested. This system provides an accurate method for the identification of the location of inhomogeneity along the length of a sheathed thermocouple and for the quantification of the inhomogeneity. This paper will discuss the apparatus and procedure used to perform these tests and will present data showing tests performed on sheathed thermocouple probes.
    Keywords: INSTRUMENTATION AND PHOTOGRAPHY
    Type: In: International Instrumentation Symposium, 39th, Albuquerque, NM, May 2-6, 1993, Proceedings (A93-54351 24-35); p. 241-254.
    Format: text
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 3
    Publication Date: 2011-08-24
    Description: An automated pressure data acquisition system for testing of pressure sensitive phosphorescent paints was designed, assembled, and tested. The purpose of the calibration system is the evaluation and selection of pressure sensitive paint chemistries that could be used to obtain global aerodynamic pressure distribution measurements. The test apparatus and setup used for pressure sensitive paint characterizations is described. The pressure calibrations, thermal sensitivity effects, and photodegradation properties are discussed.
    Keywords: INSTRUMENTATION AND PHOTOGRAPHY
    Type: In: International Instrumentation Symposium, 39th, Albuquerque, NM, May 2-6, 1993, Proceedings (A93-54351 24-35); p. 143-160.
    Format: text
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 4
    Publication Date: 2018-06-05
    Description: Temperature sensitive paints (TSP s) have recently been used to detect small temperature differences on aerodynamic model surfaces. These types of applications impose stringent performance requirements on a paint system. The TSP s must operate over a broad temperature range, must be physically robust (cannot chip or peel), must be polishable to at least the smoothness of the model surface, and must have sufficient sensitivity to detect small temperature differences. TSP coatings based on the use of metal complexes in polymer binders were developed at NASA Langley Research Center which meet most of the requirements for detection of small temperature differences under severe environmental conditions.
    Keywords: Spacecraft Propulsion and Power
    Format: application/pdf
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 5
    Publication Date: 2018-06-05
    Description: This paper presents the calibration results and uncertainty analysis of a high-precision reference pressure measurement system currently used in wind tunnels at the NASA Langley Research Center (LaRC). Sensors, calibration standards, and measurement instruments are subject to errors due to aging, drift with time, environment effects, transportation, the mathematical model, the calibration experimental design, and other factors. Errors occur at every link in the chain of measurements and data reduction from the sensor to the final computed results. At each link of the chain, bias and precision uncertainties must be separately estimated for facility use, and are combined to produce overall calibration and prediction confidence intervals for the instrument, typically at a 95% confidence level. The uncertainty analysis and calibration experimental designs used herein, based on techniques developed at LaRC, employ replicated experimental designs for efficiency, separate estimation of bias and precision uncertainties, and detection of significant parameter drift with time. Final results, including calibration confidence intervals and prediction intervals given as functions of the applied inputs, not as a fixed percentage of the full-scale value are presented. System uncertainties are propagated beginning with the initial reference pressure standard, to the calibrated instrument as a working standard in the facility. Among the several parameters that can affect the overall results are operating temperature, atmospheric pressure, humidity, and facility vibration. Effects of factors such as initial zeroing and temperature are investigated. The effects of the identified parameters on system performance and accuracy are discussed.
    Keywords: Instrumentation and Photography
    Format: application/pdf
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 6
    Publication Date: 2019-07-13
    Description: Prototype controlled-temperature enclosure containing thermoelectric devices developed to house electronically scanned array of pressure sensors. Enclosure needed because (1) temperatures of transducers in sensors must be maintained at specified set point to ensure proper operation and calibration and (2) sensors sometimes used to measure pressure in hostile environments (wind tunnels in original application) that are hotter or colder than set point. Thus, depending on temperature of pressure-measurement environment, thermoelectric devices in enclosure used to heat or cool transducers to keep them at set point.
    Keywords: PHYSICAL SCIENCES
    Type: LAR-15389 , NASA Tech Briefs (ISSN 0145-319X); 19; 8; P. 46
    Format: text
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 7
    Publication Date: 2019-07-13
    Description: The relative humidity of the atmosphere in the encasements containing the U.S. Constitution Pages 1 and 4, the Declaration of Independence, and the Bill of Rights was measured to be in the range of 55% to 61%. This value is significantly higher than the presumed relative humidity between 25 to 35 %, but is consistent with the measured samples extracted from Pages 2 and 3 of the U.S. Constitution. The cooling/condensation measurement technique used at NARA on July 23, 2001, and described in this paper to measure the water vapor content of the atmosphere in the hermetically sealed encasements containing the U. S. Constitution, the Declaration of Independence, and the Bill of Rights, proved to be a powerful new measurement technique. The cooling/condensation technique developed at NASA LaRC and utilized at NARA has important applications in the non-invasive measurement of relative humidity in the atmospheres of sealed encasements and could become a standard measurement technique in this type of analysis.
    Keywords: Instrumentation and Photography
    Type: 2004 NCSL International Workshop and Symposium; Jul 11, 2004 - Jul 15, 2004; Salt lake City, UT; United States
    Format: application/pdf
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 8
    Publication Date: 2019-07-13
    Description: Automated apparatus reveals sources of error not revealed in calibration. Computer-controlled apparatus detects and measures Seebeck inhomogeneities in sheathed thermocouples. Measures thermocouple output voltage as function of position of probe along sharp gradient of temperature. Abnormal variations in voltage-versus-position data indicative of Seebeck inhomogeneities. Prototype for development of standard method and equipment for routine acceptance/rejection testing of sheathed thermocouples in industrial and research laboratories.
    Keywords: PHYSICAL SCIENCES
    Type: LAR-15198 , NASA Tech Briefs (ISSN 0145-319X); 19; 1; P. 59
    Format: text
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 9
    Publication Date: 2019-07-12
    Description: A technique has been devised for measuring the relative humidity levels in the protective helium/water vapor atmosphere in which the Declaration of Independence, the United States Constitution, and the Bill of Rights are encased behind glass panels on display at the National Archives in Washington, DC. The technique is noninvasive: it does not involve penetrating the encasements (thereby risking contamination or damage to the priceless documents) to acquire samples of the atmosphere. The technique could also be applied to similar glass encasements used to protect and display important documents and other precious objects in museums. The basic principle of the technique is straightforward: An encasement is maintained at its normal display or operating temperature (e.g., room temperature) while a portion of its glass front panel is chilled (see Figure 1) until condensed water droplets become visible on the inside of the panel. The relative humidity of the enclosed atmosphere can then be determined as a known function of the dew point, the temperature below which the droplets condense. Notwithstanding the straightforwardness of the basic principle, careful attention to detail is necessary to enable accurate determination of the dew point. In the initial application, the affected portion of the glass panel was cooled by contact with an aluminum plate that was cooled by a thermoelectric module, the exhaust heat of which was dissipated by a heat sink cooled by a fan. A thermocouple was used to measure the interior temperature of the aluminum plate, and six other thermocouples were used to measure the temperatures at six locations on the cooled outer surface of the glass panel (see Figure 2). Thermal grease was applied to the aluminum plate and the thermocouples to ensure close thermal contact. Power was supplied to the thermoelectric module in small increments, based on previous laboratory tests. A small flashlight and a magnifying glass were used to look for water droplets condensing on the inner surface of the glass. The temperature readings of the thermocouples were taken during cool-down and upon observing condensation. In determining the dew point, it was necessary to make a correction for the differences between the temperatures measured on the chilled outer surface of the glass and the temperature of the inner surface, where the condensation took place. The correction was derived from a laboratory test on a measurement setup that was nearly identical, except that the dew location on the inner surface was also instrumented with a thermocouple. The test showed that the temperature at the dew location on the inner surface of the glass panel was 0.9 C above the temperature determined from the measurements on the chilled outer surface of the panel.
    Keywords: Man/System Technology and Life Support
    Type: LAR-16422-1 , NASA Tech Briefs, June 2005; 6-7
    Format: application/pdf
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 10
    Publication Date: 2019-07-10
    Description: In 1951, the four pages of the US Constitution, the Letter of Transmittal, the Bill of Rights and the Declaration of Independence, collectively called the 'Charters of Freedom,' were hermetically sealed individually in glass encasements. The atmosphere in the encasements consisted of a mixture of helium with water vapor at a relative humidity between 25 and 35% at room temperature. In 1998, Margaret Kelly of the National Archives and Records Administration (NARA), contacted Dr. Joel S. Levine at NASA Langley Research Center (LaRC) to request assistance in determining the chemical composition of the atmosphere inside the encasements. Several different Langley learns were assembled to address that request. each using a different measurement technique. This report describes the method and results of one team's relative humidity measurements on encasements containing pages 1 and 4 of the US Constitution, the Bill of Rights, and the Declaration of Independence performed at NARA, College Park. Maryland, on July 23, 2001.
    Keywords: Instrumentation and Photography
    Type: NASA/TM-2002-211433 , L-18145 , NAS 1.15:211433
    Format: application/pdf
    Location Call Number Expected Availability