Liquid chromatographic (LC) type trace enrichment is coupled online with capillary gas chromatography (GC) with mass spectrometric (MS) detection for the analysis of aqueous samples. A volume of 1-10 ml of an aqueous sample is preconcentrated on a trace-enrichment column packed with a polymeric stationary phase. After cleanup with HPLC-grade water the precolumn is dried with nitrogen and subsequently desorbed with ethyl acetate. A fraction of 60 μl is introduced on-line into a diphenyltetramethyldisilazane-deactivated retention gap under partially concurrent solvent evaporation conditions and using an early solvent vapor exit. The analytes are separated and detected by means of GC-MS. The potential of the LC-GC-MS system for monitoring organic pollutants in river and drinking water is studied. Target analysis is carried out with atrazine and simazine as model compounds; the detection limits achieved under full-scan and multiple ion detection conditions are 30 pg and 5 pg, respectively. Identification of unknown compounds (non-target analysis), is demonstrated using a river water sample spiked with 168 pollutants varying in polarity and volatility.