ISSN:
0934-0866
Keywords:
Chemistry
;
Industrial Chemistry and Chemical Engineering
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Mechanical Engineering, Materials Science, Production Engineering, Mining and Metallurgy, Traffic Engineering, Precision Mechanics
,
Process Engineering, Biotechnology, Nutrition Technology
Notes:
Particles comparable in size to or larger than the measurement volume need extra consideration when measured by a phase-Doppler system. The phase of the Doppler burst received when such particles traverse the measurement volume depends not only on the size of the particle but also on its trajectory, since the particle is not uniformly illuminated. This paper presents a strategy for securing correct measurements even under such conditions, taking advantage of the three-detector receiving optics of the Dantec Particle Dynamics Analyzer. The effectiveness of the approach is demonstrated for sizing drops in liquid-gas and liquid-liquid two-phase flows: water drops in air, water drops in FC72 and FC72 drops in water. The combination of water and FC72 is also of interest because the relative refractive index is close to unity. Measurements of drops size were made on a monodisperse stream of drops about 2 mm in diameter, i.e. substantially larger than the measurement volume, and polydisperse distributions of drops ranging in diameter from below 0.2 mm to about 1 mm.
Additional Material:
20 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/ppsc.19940110109
Permalink