Description:
The use of a polymeric Torlon (polyamide–imide) gasket material in a Paris–Edinburgh pressure cell forin situhigh-pressure X-ray scattering measurements is demonstrated. The relatively low bulk modulus of the gasket allows for fine control of the sample pressure over the range 0.01–0.42 GPa. The quality of the data obtained in this way is suitable for Bragg and pair distribution function analysis.
Print ISSN:
0021-8898
Electronic ISSN:
1600-5767
Topics:
Geosciences
,
Physics