ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
The development of an electron cyclotron resonance ion source for the heavy ion medical accelerator in Chiba (HIMAC) injector is reported. The HIMAC is a heavy ion medical accelerator for cancer therapy. The electron cyclotron resonance (ECR) ion source is expected to provide a long lifetime, easy operation, and easy maintenance for medical use. The NIRS-ECR ion source has a single closed ECR stage, and a microwave frequency of 10 GHz is applied. Under the present performance, the output electrical currents of the ions are 2500 eμA for He1+, 300 eμA for C2+, 480 eμA for Ne3+, and 110 eμA for Ar6+. Stability of the intensity is better than 2%. The transmission efficiency through a low-energy beam-transport line with an acceptance of 200 πmm mrad is more than 70%; the typical 50% and 90% emittances of the injection beam with 8 keV/u are 20 and 80 πmm mrad, respectively. These performances satisfy the requirements for radiotherapy.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1145071
Permalink