ALBERT

All Library Books, journals and Electronic Records Telegrafenberg

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
Filter
  • Polymer and Materials Science  (2)
Collection
Keywords
Publisher
Years
  • 1
    Electronic Resource
    Electronic Resource
    Hoboken, NJ : Wiley-Blackwell
    Journal of Biomedical Materials Research 14 (1980), S. 31-40 
    ISSN: 0021-9304
    Keywords: Chemistry ; Polymer and Materials Science
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Medicine , Technology
    Notes: Metallurgical observations of phenomena which influence the clinically successful performance of permanent implants and implant materials are reported. The effect of casting voids in cast cobalt chromium femoral stems of total hip replacements is discussed first. Pitting corrosion occurred in a retrieved stainless-steel implant type AISI 316 containing 2.7% Mo. The fractographic pattern of this device retrieved for fatigue failure shows typical characteristics of corrosion-enhanced fatigue.
    Additional Material: 3 Ill.
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
  • 2
    Electronic Resource
    Electronic Resource
    Hoboken, NJ : Wiley-Blackwell
    Journal of Biomedical Materials Research 9 (1975), S. 199-219 
    ISSN: 0021-9304
    Keywords: Chemistry ; Polymer and Materials Science
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Medicine , Technology
    Notes: A survey is given of the mechanical failures of the femoral component of Charnley and Charnley-Müller type total hip prostheses. Fractographic analysis reveals that all the prostheses have broken by metal fatigue.A review of the clinical data has been combined with a metallurgical and a mechanical study in order to assess the reasons of the mechanical failures.The following conclusions have been reached. The prosthesis should be inserted in varus position with regard to the axis of the intramedullary canal, and the prosthesis should be made of an alloy with appreciably higher yield stress and fatigue strength than the alloys of the investigated broken Charnley and Charnley-Müller type prostheses.
    Additional Material: 13 Ill.
    Type of Medium: Electronic Resource