Description:
Thermal detectors for the infrared, such as thermocouples and bolometers, are limited in their ultimate sensitivity predominantly by Johnson noise rather than temperature noise. Low noise figures are hard to achieve since Johnson noise preponderates temperature noise, which is the only essential noise for thermal detectors. The dielectric constants of some materials are sufficiently temperature dependent to make a new type of bolometer feasible. The basic theory of a dielectric bolometer, as shown here, promises noise figures below 3 decibels even at chopper frequencies well above the 1/tau value of the detector. Ferroelectrics such as barium-strontium titanate and others seem to be well suited for radiation-cooled dielectric bolometers.