Publication Date:
2011-08-16
Description:
A comparison is made of the line source (LS) method and the differential line source (DLS) method of measuring thermal conductivity of particulate materials in vacuum. The DLS method requires more instrumentation in the measuring circuitry (an additional amplifier and a differentiating circuit), but since it does not require a stable temperature to initiate a test, it does not need a sample temperature control system. DLS tests can be taken as the temperature in the samples is rising from liquid nitrogen temperature to room temperature. This eliminates the practice of extrapolating thermal conductivity over this large temperature range. Also, the advantages of reduced test time, data reduction time, and small sample temperature rise enable the experimenter to take about 7-12 DLS tests in the time of 2 LS tests. Test data from the two methods agree very well.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Review of Scientific Instruments; 44; Aug. 197