Publication Date:
2019-07-13
Description:
We have measured the temperature coefficient of the output of several ratio transformers at ratios near 0.500,000 using an ac bridge and a dual-phase, lock-in amplifier. The two orthogonal output components were each resolved to +/- ppb of the bridge drive signal. The results for three commercial ratio transformers between 20 and 50 C range from 0.5 to 100 ppb/K for the signal component in phase with the bridge drive, and from 4 to 300 ppb/K for the quadrature component.
Keywords:
ELECTRONICS AND ELECTRICAL ENGINEERING
Type:
Review of Scientific Instruments (ISSN 0034-6748); 64; 3; p. 756-759.
Format:
text
Permalink