ISSN:
0049-8246
Keywords:
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Physics
Notes:
The concept, design, characterization and operation of an apparatus for low-energy x-ray fluorescence are described. This system utilizes commercially available components in a novel arrangement. An Al K x-ray source, an ultra-thin-windowed Si(Li) detector and an ion gun for sputter cleaning were combined to perform analyses for elements with atomic number Z ≥ 6. Pairs of slit collimators placed between the specimen and the x-ray detector defined the area analyzed. The system configuration permitted analyses over areas with a variety of dimensions (0.4-22 mm). Distinction between surface and bulk signals was possible using the sputter-cleaning capability.
Additional Material:
11 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/xrs.1300190406
Permalink