    ISSN: 1612-1112
    Keywords: Column liquid chromatography ; Gel permeation ; Human serum IgG ; Fluorescent immune complex ; Fluoresence detection
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary A high-performance liquid gel-permeation chromatographic method is described for the determination of human serum immunoglobulin G (IgG) by separating the fluorescent immuno complex from the free fluorescence-labeled antibody. Fluorescence-labeled antibody used in this study was fluorescein isothiocyanate (FITC)-labeled Fab fragment goat anti-human IgG (anti-IgG Fab). Immuno complexes and antibody of different molecular sizes can be separated. FITC-labeled anti-IgG Fab was added to the serum and the mixture is passed through the column. An immuno complex separates as well-delineated peak in the column void volume, and was measured by the fluorescence of the column eluate (Ex=490nm, Em=520nm). The total analysis time for a serum sample was approximately 15min. The minimum detection limit was 25 mg/dl. The relative standard deviation was below 2% (peak area). The results of the HPL-GPC analysis correlate well with those obtained by laser nephelometric assay (r=0.992).