Publication Date:
2019-06-27
Description:
Laboratory calibrations of more than a hundred electrochemical concentration cell (ECC) ozonesondes were determined relative to UV-photometry. The average intercept and slope, 0 plus or minus 5 nb and 0.96 plus or minus 0.06, respectively, indicate reasonable agreement with UV photometry, but with considerable variation from one ECC ozonesonde to another. The time required to reach 85% of the final reaction to a step-change in ozone concentration was found to average 51 seconds. Application of the individual calibrations to 20 sets of 1976 flight data reduced the average of the differences between ozonesonde and Dobson spectrophotometric measurements of total ozone from 3.9 to 1.3%. A similar treatment of a set of 10 1977 flight records improved the average ECC-Dobson agreement from -8.5 to -1.4%. Although systematic differences were reduced, no significant effect on the random variations was evident.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
NASA-CR-155561
Format:
application/pdf
Permalink