Description:
This letter reports the detection and resolution of the characteristic X-ray of oxygen at 523 eV and other ultrasoft X-rays (photons energy less than 1 keV) using radiation detectors fabricated from the compound semi-insulator mercuric iodide (HgI2). These detectors are capable of operation at room ambient but in these experiments were slightly cooled using a Peltier element to 0 C. A pulsed light feedback preamplifier with a Peltier element cooled (to -30 deg) first stage field-effect transistor was used to amplify signals from the detector. Overall system noise level was 185 eV (full width at half-maximum) limited by the temperature of the first stage field-effect transistor. With optimal cooling of this element the characteristic X-ray of carbon at 282 eV should be measurable. These results would seem to be important in measurement of biological samples in electron column instruments.
Keywords:
ATOMIC AND MOLECULAR PHYSICS
Type:
Applied Physics Letters (ISSN 0003-6951); 46; 606