Publication Date:
2016-09-30
Description:
The note presents the first plasma density measurements collected by a novel microwave interferometer in a compact Electron Cyclotron Resonance Ion Sources (ECRIS). The developed K-band (18.5 ÷ 26.5 GHz) microwave interferometry, based on the Frequency-Modulated Continuous-Wave method, has been able to discriminate the plasma signal from the spurious components due to the reflections at the plasma chamber walls, when working in the extreme unfavorable condition λ p ≃ L p ≃ L c ( λ p , L p , and L c being the probing signal wavelength, the plasma dimension and the plasma chamber length, respectively). The note describes the experimental procedure when probing a high density plasma ( n e 〉 1 ⋅ 10 18 cm −3 ) produced by an ECRIS prototype operating at 3.75 GHz.