ISSN:
0935-6304
Keywords:
On-line extraction
;
Large volume introduction
;
Capillary gas chromatography
;
Atomic emission detection
;
Solvent venting
;
Pesticides
;
Ground water
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
The on-line coupling of a liquid-liquid extraction system with capillary gas chromatography using atomic emission detection (GC-AED) has been studied. The required large volumes of about 100 μl of an iso-octane solution can be introduced into the GC-AED system by using the AED solvent vent and a solvent vapor exit in front of the capillary analytical column. Test solutions containing several pesticides were detected using the carbon, chlorine, nitrogen and sulfur channels. Analyte detectability (in concentration units) was improved significantly and low concentractions of the test compounds could be determined (1-5 ng/ml). Aqueous samples were on-line extracted and analyzed. The precision of the large-volume injection itself as well as the total extraction-GC-AED system was satisfactory (RSD of ca. 2 and 4%, respectively). As a real-life application, several ground water samples were screened.
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240180109
Permalink