ISSN:
1612-1112
Keywords:
Column liquid chromatography
;
Anion exchange
;
Indirect photometric and direct conductivity detection
;
Naphthalenesulfonate
;
Inorganic anions
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Summary Indirect photometric and unsuppressed direct conductivity detection modes are examined using naphthalene mono-, di-, and tri-sulfonate as mobile phases for the separation of several anions such as F−, Cl−, NO2 −, Br−, NO3 −, SO4=,I−, and SCN− using a commercial anion exchange column. With all three mobile phases, conductivity detection shows better sensitivities and detection limits than indirect photometry. Conductivity detection is 5 to 16 times more sensitive than indirect photometry for all analytes. Detection limits achieved using these mobile phases are, for example, 0.04 ng and 0.1 ng for chloride ion with conductivity and indirect photometry, respectively. Both detection modes give wide linear ranges extending from at least 100 ppm to the detection limit of each anion which is generally about 0.02 ppm. Sulfur oxide anions such as dithionate and tetrathionate are separated using flow programming with naphthalenetrisulfonate as the mobile phase in less than 20 minutes. With both detection modes, desired chromatographic performance of these three eluents is achieved without pH adjustment of the mobile phase.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02276846
Permalink