ISSN:
0935-6304
Keywords:
Gas chromatography
;
Fused silica capillaries
;
Automated cold on-column injection
;
High oven temperature
;
Lipid analysis
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Modification of the Carlo-Erba cold on-column injector for (automated) analysis of high molecular compounds at high oven temperature is described. The secondary cooling tube of the cold on-column injector is replaced by a lengthened tube through which a high air flow is directed. The injection site is maintained at 65-70 °C while the oven is at high temperature (≷ 300 °C). For automated injection, a short deactivated precolumn of 22 to 30 cm × 0.53 mm i.d. is coupled to the analytical column via a butt connector with make-up gas supply. For a triglyceride mixture, automatically injected at 300 °C, the mean % deviation for all peak areas was 1.8% and the mean % deviation for all retention times was 0.09 % for five consecutive runs.
Additional Material:
9 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240100510