Publication Date:
2019-06-28
Description:
The possibility of rejecting the internal beta-decay background in coaxial germanium detectors by distinguishing between the multi-site energy losses characteristic of photons and the single-site energy losses of electrons in the range 0.2 - 2 MeV is examined. The photon transport was modeled with a Monte Carlo routine. Background rejection by both multiple segmentation and pulse shape discrimination techniques is investigated. The efficiency of a six 1 cm-thick segment coaxial detector operating in coincidence mode alone is compared to that of a two-segment (1 cm and 5 cm) detector employing both front-rear coincidence and PSD in the rear segment to isolate photon events. Both techniques can provide at least 95 percent rejection of single-site events while accepting at least 80 percent of the multi-site events above 500 keV.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
IEEE Transactions on Nuclear Science (ISSN 0018-9499); NS-31; 367-371
Format:
text
Permalink