Publication Date:
2011-08-24
Description:
An integrated-optic Mach-Zehnder interferometer is used as a Fourier transform spectrometer to analyze the input and output spectra of a temperature-sensing thin-film etalon. This type of spectrometer has an advantage over conventional grating spectrometers because it is better suited for use with time-division-multiplexed sensor networks. In addition, this spectrometer has the potential for low cost due to its use of a component that could be manufactured in large quantities for the optical communications industry.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
In: Integrated optics and microstructures; Proceedings of the Meeting, Boston, MA, Sept. 8, 9, 1992 (A93-49456 21-35); p. 127-136.
Format: