    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 68 (1997), S. 3728-3737 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: In this article we describe the design and the operation of an original, high transmission, electrostatic "double toroidal" electron energy analyzer. The double toroidal analyzer allows the high resolution and high luminosity simultaneous measurement of the kinetic energy, and angular distribution of electrons, using a two-dimensional position sensitive detector. The exact shape of the electrodes is deduced from both analytical and numerical electron trajectory calculations. The electron detector is based on a charge analysis and optimized to attain a 100 kHz counting rate. The actual performances of the analyzer are illustrated with spectra obtained after resonant Auger decay of N2O excited around the nitrogen K shell (hν=401 eV), and of Kr after 3d5/2→5p excitation at hν=91.2 eV. A "étendue" of 15% of the pass energy, as well as a resolving power (Ep/δE) of 100were measured. © 1997 American Institute of Physics.