Abstract

Starting from a rigorous formulation for the admittance of an open ended coaxial probe, the authors identify the simplifying assumptions that are introduced to obtain more practical solutions. The predictions of the models are compared to each other and to experimental data obtained on an independently calibrated experimental set-up. The advantages and limitations of a model are determined by the theoretical derivation and the numerical solution. The model identified as most suitable for the measurement of the dielectric properties of biological materials is used to determine the relative permittivity and conductivity of muscle and skull bone in the frequency range 1 MHz to 20 GHz.