Abstract
The noise characteristic of available laser sources limits the sensitivity of saturated absorption spectroscopy. Time averaging is one method to improve the signal-to-noise ratio. A computer controlled dye-laser spectrometer is described which has the capability to scan several times any selected frequency range with an absolute accuracy of ±200 kHz. The sensitivity of this system is demonstrated by measuring the isotope shift of the low abundant38Ar and by detecting a weak83Kr-hyperfine component.
Similar content being viewed by others
References
H. Gerhardt, A. Timmermann: Opt. Commun.21, 343 (1977)
H. Gerhardt, E. Matthias, F. Schneider, A. Timmermann: Z. Physik A288, 327 (1978)
W.G. Schweitzer, Jr., E.G. Kessler, Jr., R.D. Deslattes, H.P. Layer, J.R. Whetstone: Appl. Opt.12, 2927 (1973)
R.J. Wallard: J. Phys. E5, 926 (1972)
G.R. Hanes, K.M. Baird, J. De Remigis: Appl. Opt.12, 1600 (1973)
V. Henc-Bartolic, D. Soldo, A. Persin, M. Dusi: J. Quant. Spectrosc. Radiat. Transfer23, 237 (1980)
C. Fourrier, G. Lenperierre, J.M. Poitevin: Phys. Lett.74A, 313 (1979)