Abstract The comparison of an ambient temperature mercuric iodide spectrometer (HgI2) with a conventional cooled Si(Li) spectrometer is presented for the determination of niobium in Venezuelan laterites in the laboratory. The energy resolution of the HgI2 spectrometer was only about 590 eV (FWHM) at 5.9 keV compared to about 150 eV for the conventional Si(Li) spectrometer. The efficiency was about 5 times lower in the region of the NbK α peak as compared to the Si(Li) detector. Even though its energy resolution and efficiency is much poorer than the conventional Si(Li), the results it produced for the determination of niobium were acceptable for use in the exploration and exploitation of a niobium-rich deposit such as Cerro Impacto, Estado Bolívar, Venezuela.