ISSN:
1573-9686
Keywords:
Hearing testing
;
Adaptive noise canceling
;
Adaptive Wiener filtering
;
Otoacoustic emissions
Source:
Springer Online Journal Archives 1860-2000
Topics:
Medicine
,
Technology
Notes:
Abstract An innovative acoustic noise canceling method using adaptive Wiener filtering (AWF) was developed for improved acquisition of distortion product otoacoustic emissions (DPOAEs). The system used one microphone placed in the test ear for the primary signal. Noise reference signals were obtained from three different sources: (a) pre-stimulus response from the test ear microphone, (b) post-stimulus response from a microphone placed near the head of the subject and (c) post-stimulus response obtained from a microphone placed in the subject’s nontest ear. In order to improve spectral estimation, block averaging of a different number of single sweep responses was used. DPOAE data were obtained from 11 ears of healthy newborns in a well-baby nursery of a hospital under typical noise conditions. Simultaneously obtained recordings from all three microphones were digitized, stored and processed off-line to evaluate the effects of AWF with respect to DPOAE detection and signal-to-noise ratio (SNR) improvement. Results show that compared to standard DPOAE processing, AWF improved signal detection and improved SNR. © 1998 Biomedical Engineering Society. PAC98: 4364Jb, 4360-c, 8790+y
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1114/1.111
Permalink