We report the design and testing of a nuclear magnetic resonance (NMR) microcoil receiver apparatus, employing solenoidal microreceiver coils of dimensions of tens to hundreds of microns, using applied field of 9 T (proton resonance frequency 383 MHz). For the smallest receiver coils we attain sensitivity sufficient to observe proton NMR with signal to noise (S/N) one in a single scan applied to a ∼10 μm3 (10 fl) water sample, containing 7×1011 total proton spins. We also test the dependence of the S/N on important coil parameters, including coil composition and resistivity, turn spacing, and lead lengths. © 2001 American Institute of Physics.