ISSN:
0006-3592
Keywords:
hybridoma cultivation media
;
glucose
;
L-glutamine
;
flow-injection analysis
;
biosensors
;
Chemistry
;
Biochemistry and Biotechnology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Biology
,
Process Engineering, Biotechnology, Nutrition Technology
Notes:
A split-stream flow-injection analysis system is described for simultaneous determination of glucose and L-glutamine in serum-free hybridoma bioprocesses media. Amperometric measurement of glucose is based on anodic oxidation of hydrogen peroxide produced by immobilized glucose oxidase within a triple layer membrane of an integrated flow-through glucose-selective biosensor. Determination of L-glutamine is based on quantitating ammonium ions produced in a flow-through enzymes reactor containing immobilized glutaminase enzyme, and subsequent downstream potentiometric detection of these ions by a nonacting-based ion-selective polymer membrane electrode. Endogenous potassium and ammonium ion interference in the L-glutamine determination are eliminated by using a novel in-line tubular cation-exchange membrane unit to exchange these interferent species for cations undetectable by the membrane electrode. The first generation split-steam flow-injections system can assay 12 samples/h using direct injections of 50 μL of media samples, with linear responses to glucose in the range of 0.03 to 30mM, and log-linear response to L-glutamine from 0.1 to 10 mM. © 1993 Wiley & Sons, Inc.
Additional Material:
5 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/bit.260411007
Permalink