ISSN:
1572-8943
Schlagwort(e):
Knudsen effusion method
;
pressure DSC
;
steel ball method
;
vapor pressure determination
Quelle:
Springer Online Journal Archives 1860-2000
Thema:
Chemie und Pharmazie
Notizen:
Abstract A new pressure DSC module (Mettler DSC27HP) and its abilities for vapor pressure determination in the range of subambient pressure to 7 MPa are presented. To compare the new to an established method, vapor pressures of caffeine, naphthalene and o-phenacetin have been determined both by pressure DSC and the Knudsen effusion cell method. These results, including the derived heats of evaporation and heats of sublimation, are compared to literature values.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1023/A:1010133106907