    Publication Date: 2019-08-28
    Description: A fine-wire thermocouple probe was used to determine freestream stagnation temperatures in hypersonic flows. Data were gathered in a N2 blowdown wind tunnel with runtimes of 1-5 s. Tests were made at supply pressures between 30 and 1400 atm and supply temperatures between 700 and 1900 K, with Mach numbers of 14 to 16. An iterative procedure requiring thermocouple data, pilot pressure measurements, and supply conditions was used to determine test cell stagnation temperatures. Probe conduction and radiation losses, as well as real gas behavior of N2, were accounted for during analysis. Temperature measurement error was found to be 5 to 10 percent. A correlation was drawn between thermocouple diameter Reynolds number and temperature recovery ratio. Transient probe behavior was studied and was found to be adequate in temperature gradients up to 1000 K/s.
    Keywords: INSTRUMENTATION AND PHOTOGRAPHY
    Type: ICIASF ''91 - International Congress on Instrumentation in Aerospace Simulation Facilities; Oct 27, 1991 - Oct 31, 1991; Rockville, MD; United States