ISSN:
0741-0581
Schlagwort(e):
High voltage electron microscopy
;
Energy loss spectroscopy
;
Life and Medical Sciences
;
Cell & Developmental Biology
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Allgemeine Naturwissenschaft
Notizen:
This paper describes work on the counting of electrons in high voltage electron microscopy (HVEM). A reliable counting method is needed in order to perform quantitative experiments. A scintillator-photomultiplier device has been adapted to these needs. It has been used here, in the field of electron energy loss spectroscopy. It permits the recording of the totality of the spectrum from the no-loss peak to the inner shell excitation characteristic signals because of a dynamic equal to at least 7. A simple mounting has been made to eliminate the artifacts caused by X-ray pulses. It is shown that HVEM is favourable to the counting of electrons, and some applications are reported.
Zusätzliches Material:
17 Ill.
Materialart:
Digitale Medien
URL:
http://dx.doi.org/10.1002/jemt.1060030112
Permalink