ISSN:
1432-1130
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Abstract A method has been proposed for the determination of mercury by cold vapor generation graphite furnace atomic absorption spectrometry (CV-GFAAS) with Pd-Rh as coating and chemical modifier. The trapping efficiency for mercury with Pd-Rh was higher than with Pd alone. The characteristic mass of the method, which gives an integrated absorbance of 0.0044 s, was found to be 55 pg and the absolute detection limit (3 σ) of 37 pg was obtained with the proposed modifier. The method was successfully applied to the determination of mercury in standard reference water samples, wastewater samples and cosmetics with a recovery range of 95–104%.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/s002160051203