The purpose of this pilot study was to evaluate surface-induced mineralization (SIM) as a potential technique to apply ceramic coatings to metal orthopaedic implants. Cylindrical titanium porous-coated implants were either coated by SIM or plasma-spray (PLS) techniques with calcium phosphate, or left uncoated (CTL). The implants were bilaterally implanted into the intramedullary canal of the proximal femur of 24 adult New Zealand white rabbits segregated into the following groups: PLS/CTL, SIM/CTL, and SIM/PLS. After 6 weeks in vivo, biomechanical and histologic evaluations were completed. Biomechanically, SIM had consistently greater mechanical interlock than PLS implants. However, CTL implants had greater mechanical interlock than both PLS and SIM. The small sample size prevented statistical evaluation and definitive biomechanical conclusions. Histologically, SIM and PLS had significantly greater ingrowth than CTL implants (p 〈 0.05). The SIM coating technique produced similar ingrowth characteristics as standard PLS coatings, yet may prevent osteolysis by providing a stronger, more reliable, covalent bond between the ceramic and metal. © 1997 John Wiley & Sons, Inc.